National Tsing Hua University Institutional Repository:Performance evaluation of graphite pancake ionization chamber by comparing the absorbed dose to water calibration
English  |  正體中文  |  简体中文  |  全文笔数/总笔数 : 54367/62174 (87%)
造访人次 : 14675560      在线人数 : 204
RC Version 6.0 © Powered By DSPACE, MIT. Enhanced by NTHU Library IR team.
搜寻范围 查询小技巧:
  • 您可在西文检索词汇前后加上"双引号",以获取较精准的检索结果
  • 若欲以作者姓名搜寻,建议至进阶搜寻限定作者字段,可获得较完整数据
  • 进阶搜寻
    National Tsing Hua University Institutional Repository > 原子科學院  > 生醫工程與環境科學系 > 期刊論文 >  Performance evaluation of graphite pancake ionization chamber by comparing the absorbed dose to water calibration


    jsp.display-item.identifier=請使用永久網址來引用或連結此文件: http://nthur.lib.nthu.edu.tw/dspace/handle/987654321/61824


    题名: Performance evaluation of graphite pancake ionization chamber by comparing the absorbed dose to water calibration
    作者: Lee JH;Lin CY;Wang CF;Butler DJ;Webb DV;Yeh CY;Hwang WS
    教師: 王竹方
    日期: 2006
    出版者: Elsevier
    關聯: Radiation Measurements, Volume 41, Issue 3, Pages 306-316
    关键词: graphite pancake ionization chamber
    absorbed dose to water
    TG-21
    TG-51
    摘要: This paper describes modifications to an original design, correction factors and uncertainty evaluations for a graphite pancake ionization chamber constructed at the Institute of Nuclear Energy Research (INER, Taiwan). A bilateral comparison of the absorbed dose to water standards for Co-60 using transfer chambers was performed to verify the graphite chamber experimental accuracy and measurement consistency. The comparison results showed a satisfactory agreement in the measurements within the combined standard uncertainties (k = 1). This paper also compares the absorbed dose to water calibration in medical accelerator photon beams traceable to INER standards following the recommendations given in the AAPM TG-21 and TG-51 dosimetry protocols. For all types of linear accelerators and cylindrical chambers at 13 participating hospitals in Taiwan, the TG-51/TG-21 dose ratios were the same within +/- 1.5%, less than the combined uncertainty, irrespective of the chamber make and model for each photon included here. A quality assurance guide for institutions switching from the TG-21 to TG-51 protocol was suggested based on the comparison results for the two dosimetry protocols.
    URI: http://www.elsevier.com/
    http://nthur.lib.nthu.edu.tw/dspace/handle/987654321/61824
    显示于类别:[生醫工程與環境科學系] 期刊論文

    文件中的档案:

    档案 描述 大小格式浏览次数
    index.html0KbHTML677检视/开启


    在NTHUR中所有的資料項目都受到原著作權保護,僅提供學術研究及教育使用,敬請尊重著作權人之權益。若須利用於商業或營利,請先取得著作權人授權。
    若發現本網站收錄之內容有侵害著作權人權益之情事,請權利人通知本網站管理者(smluo@lib.nthu.edu.tw),管理者將立即採取移除該內容等補救措施。

    SFX Query

    與系統管理員聯絡

    DSpace Software Copyright © 2002-2004  MIT &  Hewlett-Packard  /   Enhanced by   NTU Library IR team Copyright ©   - 回馈