English  |  正體中文  |  简体中文  |  Items with full text/Total items : 54367/62174 (87%)
Visitors : 10197588      Online Users : 132
RC Version 6.0 © Powered By DSPACE, MIT. Enhanced by NTHU Library IR team.
Scope Tips:
  • please add "double quotation mark" for query phrases to get precise results
  • please goto advance search for comprehansive author search
  • Adv. Search
    HomeLoginUploadHelpAboutAdminister Goto mobile version
    National Tsing Hua University Institutional Repository > 原子科學院  > 生醫工程與環境科學系 > 期刊論文 >  Dosimertry characteristics and performance comparisons: Environmental radiophotoluminescent glass dosemeters versus thermoluminescent dosemeters


    Please use this identifier to cite or link to this item: http://nthur.lib.nthu.edu.tw/dspace/handle/987654321/61832


    Title: Dosimertry characteristics and performance comparisons: Environmental radiophotoluminescent glass dosemeters versus thermoluminescent dosemeters
    Authors: Lee JH;Lin MS;Hsu SM;Chen IJ;Chen WL;Wang CF
    教師: 王竹方
    Date: 2009
    Publisher: Elsevier
    Relation: RADIATION MEASUREMENTS, Volume: 44, Issue: 1, Pages: 86-91
    Keywords: Radiophotoluminescent glass dosemeter
    Thermoluminescent dosemeter
    Cumulative dose
    Fading
    Performance test
    Abstract: This paper describes the dosimetry characteristic comparison of environmental radiophotoluminescent glass dosemeter (RPLGD) and thermoluminescent dosemeter (TLD) systems employed at the Institute of Nuclear Energy Research (INER, Taiwan). The luminescence centers of TLD disappeared by reading process, and repetition of measurement is impossible. RPLGDs can be repeatedly read and keep the luminescence centers for a long time. The RPLGD fading is about 1% within 30 days after being exposed to ionizing radiation. Environmental cumulative doses measured by a RPLGD and a TLD were compared at the same monitoring points of INER during two years. The sensitivity of TLD decreased with gradual loss due to fading at higher temperatures. The difference between results obtained by RPLGD and TLD is discussed mainly with respect to the dependence on the ambient temperature. In addition, this research also refers to the American National Standard Institute (ANSI) Draft Standard N13.29 (1996) to organize the environmental dosimetry performance tests of the INER's RPLGDs and the TLDs. The results mean that both kinds of dosemeters can meet the performance test criteria. An accreditation procedure for the institutes which provide the environmental dosimetry services in Taiwan was suggested based on the comparison results of these two dosimetry systems.
    URI: http://www.elsevier.com/
    http://nthur.lib.nthu.edu.tw/dspace/handle/987654321/61832
    Appears in Collections:[生醫工程與環境科學系] 期刊論文

    Files in This Item:

    File Description SizeFormat
    index.html0KbHTML563View/Open


    在NTHUR中所有的資料項目都受到原著作權保護,僅提供學術研究及教育使用,敬請尊重著作權人之權益。若須利用於商業或營利,請先取得著作權人授權。
    若發現本網站收錄之內容有侵害著作權人權益之情事,請權利人通知本網站管理者(smluo@lib.nthu.edu.tw),管理者將立即採取移除該內容等補救措施。

    SFX Query

    與系統管理員聯絡

    DSpace Software Copyright © 2002-2004  MIT &  Hewlett-Packard  /   Enhanced by   NTU Library IR team Copyright ©   - Feedback