    National Tsing Hua University Institutional Repository > 原子科學院  > 生醫工程與環境科學系 > 期刊論文 >  Dosimertry characteristics and performance comparisons: Environmental radiophotoluminescent glass dosemeters versus thermoluminescent dosemeters

    Please use this identifier to cite or link to this item: http://nthur.lib.nthu.edu.tw/dspace/handle/987654321/61832

    Title: Dosimertry characteristics and performance comparisons: Environmental radiophotoluminescent glass dosemeters versus thermoluminescent dosemeters
    Authors: Lee JH;Lin MS;Hsu SM;Chen IJ;Chen WL;Wang CF
    教師: 王竹方
    Date: 2009
    Publisher: Elsevier
    Relation: RADIATION MEASUREMENTS, Volume: 44, Issue: 1, Pages: 86-91
    Keywords: Radiophotoluminescent glass dosemeter
    Thermoluminescent dosemeter
    Cumulative dose
    Performance test
    Abstract: This paper describes the dosimetry characteristic comparison of environmental radiophotoluminescent glass dosemeter (RPLGD) and thermoluminescent dosemeter (TLD) systems employed at the Institute of Nuclear Energy Research (INER, Taiwan). The luminescence centers of TLD disappeared by reading process, and repetition of measurement is impossible. RPLGDs can be repeatedly read and keep the luminescence centers for a long time. The RPLGD fading is about 1% within 30 days after being exposed to ionizing radiation. Environmental cumulative doses measured by a RPLGD and a TLD were compared at the same monitoring points of INER during two years. The sensitivity of TLD decreased with gradual loss due to fading at higher temperatures. The difference between results obtained by RPLGD and TLD is discussed mainly with respect to the dependence on the ambient temperature. In addition, this research also refers to the American National Standard Institute (ANSI) Draft Standard N13.29 (1996) to organize the environmental dosimetry performance tests of the INER's RPLGDs and the TLDs. The results mean that both kinds of dosemeters can meet the performance test criteria. An accreditation procedure for the institutes which provide the environmental dosimetry services in Taiwan was suggested based on the comparison results of these two dosimetry systems.
    URI: http://www.elsevier.com/